The purpose of this study was to determine the influence of the reference electrode to the motor unit action potential parameters including amplitude, duration, surface area, phases and turns by reading the same recording area by monopolar and cocentric needles. A concentric needle was stuck into the subject's anterior tibialis muscle, online motor unit analysis was done by EMG-computer combinations. The sinal from the active recording electrode was shunted to another EMG-computer set for analysis of the same motor unit by adding a surface electrode as a reference electrode (monopolar needle recording pattern). After collecting 200 motor unit potentials, data analysis of five parameters were made for comparison. Significant differences in duration and surface area were found, but not in amplitude, turns and phases. Factors contributing to the results are discussed.